To continue calibration from step 6 proceed as follows:
Press [UP] to go to CAL step 7. (ACA 3mA Offset
0.12mA)
Press [STOP]
Press [START]
Using keypad enter the reading from the current meter followed by the [ENTER].
Press [STOP] to accept reading.
Press [UP] to go to CAL step 8. (ACA 3mA Full
2.500mA)
Press [STOP]
Press [START]
Using keypad enter the reading from the current meter followed by the [ENTER].
Press [STOP] to accept reading.
Press [UP] to go to CAL step 9. (ACA 40mA Offset
2.500mA)
Press [STOP]
Press [START]
Using keypad enter the reading from the current meter followed by the [ENTER].
Press [STOP] to accept reading.
Press [UP] to go to CAL step 10. (ACA 40mA Full
25mA)
Press [STOP]
Press [START]
Using keypad enter the reading from the current meter followed by the [ENTER].
Press [STOP] to accept reading.
Service & Calibration
150799 Rev A3
Page 169 of 177
www.valuetronics.com