Download Print this page

QuadTech Guardian 6100 Plus Instruction Manual page 141

Electrical safety analyzer

Advertisement

:STEP <n>
:AC
:LEVel <numeric value>
Command: :SOURce : SAFEty : STEP <n> : AC : LEVEL
This command sets the AC voltage level for the selected step.
<numeric value> is 50 to 5000 and the unit is volts.
:LEVel?
Command: :SOURce : SAFEty : STEP <n> : AC : LEVEL?
This command queries the AC voltage level for the selected step.
The return format is a numerical value: 50 ≤ value ≤ 5000 volts.
:LIMit
Interface
www.valuetronics.com
:AREPort ? (RS232 Interface only)
Command: :SOURce : SAFEty : RESult : AREPort?
This command queries if the Automatic Reporting of test results is ON or OFF.
The return format is a "1" or a "0".
:SNUMber?
Command: :SOURce : SAFEty : SNUMber?
This command queries the step number being set in memory.
:DELete
Command: :SOURce : SAFEty : STEP : DELete
This command clears all setting values in selected step to the initial value.
<n> denotes the step number. Range: 1 ≤ n ≤ 99.
:SET?
Command: :SOURce : SAFEty : STEP : SET?
This command queries all setting values in selected step.
:MODE?
Command: :SOURce : SAFEty : STEP <n> : MODE?
This command queries the test Mode of the selected step.
The return format is: GB⏐AC⏐DC⏐IR⏐OSC⏐PA⏐LC.
:HIGH <numeric value>
Command: :SOURce : SAFEty : STEP <n> : AC : LIMit : HIGH
This command sets the high current limit for the selected step.
<numeric value> is 0.000001 to 0.030 and the unit is amps.
:HIGH?
Command: :SOURce : SAFEty : STEP <n> : AC : LIMit : HIGH?
This command queries the high current limit for the selected step.
:LOW <numeric value>
Command: :SOURce : SAFEty : STEP <n> : AC : LIMit : LOW
This command sets the low current limit for the selected step.
<numeric value> is 0 to 0.030 and the unit is amps. 0 = OFF
:LOW?
Command: :SOURce : SAFEty : STEP <n> : AC : LIMit : LOW?
This command queries the low current limit for the selected step.