:TIME
:CHANnel (@ <Box#> (<channel list>))
Page 142 of 177
www.valuetronics.com
:ARC <numeric value>
Command: :SOURce : SAFEty : STEP <n> : AC : LIMit : ARC
This command sets the arc current limit for the selected step.
<numeric value> is 0 to 0.030 and the unit is amps. 0 = OFF
:ARC?
Command: :SOURce : SAFEty : STEP <n> : AC : LIMit : ARC?
This command queries the arc current limit for the selected step.
:ARC:FILTER <numeric value>
Command: :SOURce : SAFEty : STEP <n> : AC : LIMit : ARC FILTER
This command sets the arc filter for the selected step.
<numeric value> is 23e3, 50e3, 100e3 or 230e3 and the unit is Hz.
:ARC FILTER?
Command: :SOURce : SAFEty : STEP <n> : AC : LIMit : ARC FILTER?
This command queries the arc filter for the selected step.
:RAMP <numeric value>
Command: :SOURce : SAFEty : STEP <n> : AC : TIME : RAMP
This command sets the ramp time for the selected step.
<numeric value> is 0, 0.1 to 999 and the unit is seconds. 0 = OFF
:RAMP?
Command: :SOURce : SAFEty : STEP <n> : AC : TIME : RAMP?
This command queries the ramp time for the selected step.
:TEST <numeric value>
Command: :SOURce : SAFEty : STEP <n> : AC : TIME : TEST
This command sets the test time for the selected step.
<numeric value> is 0, 0.3 to 999 and the unit is seconds. 0 = OFF
:TEST?
Command: :SOURce : SAFEty : STEP <n> : AC : TIME : TEST?
This command queries the test time for the selected step.
:HIGH <channel list>
Command: :SOURce : SAFEty : STEP <n> : AC : CHAN : HIGH (@ Box#
(<channel list>))
This command sets the high scan channels for the selected step.
:HIGH?
Command: :SOURce : SAFEty : STEP <n> : AC : CHAN : HIGH?
This command queries the high scan channels for the selected step.