Download Print this page

QuadTech Guardian 6100 Plus Instruction Manual page 10

Electrical safety analyzer

Advertisement

Insulation Resistance
Voltage:
Accuracy:
Range:
Accuracy:
Ground Bond
Output Current:
Resistance:
Page 10 of 177
www.valuetronics.com
Specifications (Continued)
50 - 1000V DC in 1V steps
± (1% of reading + 5V)
100kΩ - 50GΩ (voltage dependent)
± (5% + 5 counts) ≥ 500V, 10MΩ - 1GΩ
± (10% + 5 counts) ≥ 500V, 1GΩ - 10GΩ
± (15% + 5 counts) ≥ 500V, 10GΩ - 50GΩ
± (10% + 5 counts) < 500V, 100kΩ - 1GΩ
Voltage
1000V
Specification Range
500V
50V
0.1M
20M
2M
Range:
Regulation:
Accuracy:
Frequency:
Voltage:
Range:
Accuracy:
Resolution:
Current
30A
20A
Specification Range
12A
10A
8A
6A
4A
3A
2A
20
40 60
80
100
0.1mΩ
1G
50G
5G
1.0 to 30.0A AC, setting .01A/step
To 40A with optional transformer (G45)
± (1% of setting + 0.3A)
± (1% of reading + 3 counts)
50/60Hz, selectable
6 to 15V, programmable
0.1mΩ - 510.0m
, 4 digits, Hi Limit
± (1% of reading + 3 counts)
0.1mΩ
200
300
400
210mΩ
150799 Rev A3
Resistance
Resistance
500
510mΩ

Advertisement