2.3.1 SYSTEM
To access SYSTEM SETUP, press the [MENU] function key at the top left hand corner
of the keypad and press the numerical key [1]. Within SYSTEM SETUP there are 11
parameters: Contrast, Beeper Volume, Compensate, DC 50V AGC, Discharge Voltage,
Pass ON, Use Source, After Fail, AC Offset, LC Offset and IEEE-488.1. Refer to Figure
2-2.
To access any of these 11 parameters, press the [↑] or [↓] arrow on the keypad to move
the backlit box to the desired parameter. The status window in the bottom left-hand
corner of the display shows the range for the selected parameter. Use the function keys
or the numerical keys and [ENTER] to set a SYSTEM parameter.
The G6100 Plus can be used with an AC Source (¶2.3.1.7). In the SYSTEM menu, select USE SOURCE
= ON so that in LC Mode the POWER option SOURCE is operational. Set POWER = Source; TARGET-
Operation
www.valuetronics.com
SYSTEM SETUP
01.
Contrast
:
02.
Beeper Vol.
:
03.
Compensate
:
04.
DC 50V AGC
:
05.
Discharge - V
:
06.
Pass ON
:
07.
Use Source
:
08.
After Fail
:
09.
AC Offset
:
10.
LC Offset
:
11.
IEEE-488.1
:
1 - 31
Figure 2-2: SYSTEM SETUP
Table 2-2: System Parameters
Parameter
Contrast
Beeper Volume
HIGH, MEDIUM, LOW, OFF
Compensate
DC 50V AGC
Discharge-V
PASS On
Use Source
After Fail
CONTINUE, STOP, START
AC Offset
LC Offset
IEEE-488.1
V in the range 80-300V and TARGET-F = 50 or 60Hz.
150799 Rev A3
17
UP
HIGH
20%
ON
DOWN
3.10kV
1.0s
OFF
STOP
0.5mA
0.00mA
ON
Remote
Lock
Offset
Range
1-31
±(5-50%) or OFF
ON/OFF
0.05 – 5.1kV
0.1 – 99.9s, CONT
ON/OFF
0 – 2.5mA
0 – 2.5mA
ON/OFF
NOTE:
Error
Default
17
HIGH
20%
ON
5.1kV
CONT
OFF
STOP
0.50mA
0.00mA
OFF
Page 35 of 177