:TIME
:POWer
Interface
www.valuetronics.com
:LIMit:LOW <numeric value>
Command: :SOURce:SAFEty:STEP<n>:LC:LIMit:LOW <numeric value>
This command sets the low current limit for the selected step.
<numeric value> is 0 to 10 and the unit is milliamps.
:LIMit:LOW?
Command: :SOURce:SAFEty:STEP<n>:LC:LIMit:LOW?
Return Value: +1.000000E-02
This command queries the low current limit for the selected step.
[:TEST]<numeric value>
Command: :SOURce:SAFEty:STEP <n>:LC:TIME[:TEST] <numeric value>
This command sets the test time for the selected step.
<numeric value> is 0 to 999 and the unit is seconds.
[:TEST]?
Command: :SOURce:SAFEty:STEP <n>:LC:TIME[:TEST]?
Return Value: +6.000000E+01 (for 60 second test)
This command queries the test time for the selected step.
:MODE?
Command: :SOURce:SAFEty:STEP <n>:LC:POWer:MODE?
Return: VOLTAGE | CURRENT | VA | SOURCE
This command queries the power measurement mode for the selected step.
:VOLTage
Command: :SOURce:SAFEty:STEP <n>:LC:POWer:VOLTage
This command sets the power measurement to voltage for the selected step.
:LIMit[:HIGH] <numeric value>
Command: :SOURce:SAFEty:STEP<n>:LC:POWer:VOLTage[:LIMit][:HIGH]<nv>
This command sets the high power limit for the selected step.
<numeric value> is 0 to 300 and the unit is volts
:LIMit[:HIGH]?
Command: :SOURce:SAFEty:STEP<n>:LC: POWer:VOLTage[:LIMit][:HIGH]?
This command queries the high power limit for the selected step
The return format is a numerical value: 0 ≤ value ≤ 300V
:LIMit:LOW <numeric value>
Command: :SOURce:SAFEty:STEP<n>:LC: POWer:VOLTage[:LIMit]:LOW <nv>
This command sets the low power limit for the selected step.
<numeric value> is 0 to 300 and the unit is volts.
:LIMit:LOW?
Command: :SOURce:SAFEty:STEP<n>:LC: POWer:VOLTage[:LIMit]:LOW?
This command queries the low power limit for the selected step
The return format is a numerical value: 0 ≤ value ≤ 300V, 0 = OFF