AC Hipot Programming – continued from last page.
Press DOWN arrow [⇓]
STEP SETTING
01.
Test Step
02.
Test Mode
03.
Voltage
04.
High Limit
05.
Low Limit
06.
Arc Limit
07.
Arc Filter
08.
Test Time
09.
Ramp Time
10.
CHNL (H-L)
0 - 20 mA
Press [0][ENTER] to set arc current limit = OFF
Press DOWN arrow [⇓]
STEP SETTING
01.
Test Step
02.
Test Mode
03.
Voltage
04.
High Limit
05.
Low Limit
06.
Arc Limit
07.
Arc Filter
08.
Test Time
09.
Ramp Time
10.
CHNL (H-L)
SELECT FILTER
Press the function key [3 -230 kHz] to set arc filter frequency = 3 – 230kHz
Press DOWN arrow [⇓]
STEP SETTING
01.
Test Step
02.
Test Mode
03.
Voltage
04.
High Limit
05.
Low Limit
06.
Arc Limit
07.
Arc Filter
08.
Test Time
09.
Ramp Time
10.
CHNL (H-L)
0, 0.3 - 999 sec
Press [3][.][0][ENTER] to set test time = 3.0s
Continued on next page.
Operation
www.valuetronics.com
3
:
:
AC
1.500kV
:
:
0.500mA
OFF
:
OFF
:
3-230 kHz
:
3.0s
:
OFF
:
OFF
:
Remote
Lock
Offset
3 - 23 kHz
:
3
:
AC
3 - 50 kHz
1.500kV
:
:
0.500mA
:
OFF
3 - 100 kHz
OFF
:
3-230 kHz
:
3 - 230 kHz
3.0s
:
OFF
:
OFF
:
Remote
Lock
Offset
3
:
:
AC
1.500kV
:
:
0.500mA
OFF
:
OFF
:
3-230 kHz
:
3.0s
:
OFF
:
OFF
:
0 =CONT
Remote
Lock
Offset
150799 Rev A3
Use Numeric Keys then [ENTER]
Enter the Arc Limit in milli-amps (mA)
0 - 20 mA, 0 = OFF
Error
Use Function Keys
Select the Arc Filter Frequency in kilo-hertz (kHz)
3 -23 kHz
3 -50 kHz
3 -100 kHz
3 -230 kHz
Error
Use Numeric Keys then [ENTER]
Enter the Test Time in seconds (s)
0, 0.3 - 999 seconds; 0 = Continuous
Error
(40µs)
(20µs)
(10µs)
(4µs)
Page 63 of 177