Download Print this page

QuadTech Guardian 6100 Plus Instruction Manual page 148

Electrical safety analyzer

Advertisement

:LC
:DEVice
Command: :SOURce:SAFEty:STEP<n>:LC:DEVice
This command sets the circuit model for the selected step.
UL1950 | UL 1563 | UL544NP | UL544P | UL2601
:DEVice?
Command: :SOURce:SAFEty:STEP<n>:LC:DEVice?
This command queries the circuit model for the selected step.
:LINE
Command: :SOURce:SAFEty:STEP<n>:LC:LINE
This command sets the line input for the selected step.
NORmal | REVerse | SFNormal | SFReverse
:LINE?
Command: :SOURce:SAFEty:STEP<n>:LC:LINE?
This command queries the line input for the selected step.
:METEr
Command: :SOURce:SAFEty:STEP<n>:LC:METEr
This command sets the measurement point for the selected step.
:METEr?
Command: :SOURce:SAFEty:STEP<n>:LC:METEr?
This command queries the measurement point for the selected step.
Return: L,G | L,P | P,P
:GSWItch <boolean> 1 | 0
Command: :SOURce:SAFEty:STEP<n>:LC:GSWItch
This command sets the ground switch for the selected step.
:GSWItch?
Command: :SOURce:SAFEty:STEP<n>:LC:GSWItch?
This command queries the ground switch for the selected step.
Return: 1 | 0 1 = ON = Open; 0 = OFF = Closed
:LIMit
Page 148 of 177
www.valuetronics.com
:LIMit[:HIGH] <numeric value>
Command: :SOURce:SAFEty:STEP<n>:LC:LIMit[:HIGH] <numeric value>
This command sets the high current limit for the selected step.
<numeric value> is 0 to 10 and the unit is milliamps
:LIMit[:HIGH]?
Command: :SOURce:SAFEty:STEP<n>:LC:LIMit[:HIGH]?
Return Value: +1.000000E-03
This command queries the high current limit for the selected step