Offset – continued
The following formulas apply to the offset function:
For DC offset:
For AC offset:
For LC offset:
NOTE
When in LC Mode, the following must hold true: Offset Leakage + High Limit < Measurement Range
For example, when the High Limt is 0.50mA and the instrument is in the 5999 mA range, if Offset current
> 0.1mA, any attempt to perform an Offset will result in HIGH FAILURE.
2.7.2 GET Cs (OSC)
Refer to ¶ 2.4.2 for a full description of the Open/Short Circuit detection mode. In
MAIN MENU, press [TEST] to enter Test Menu. Run OFFSET first (open circuit) and
then connect the 'known good' DUT. Press the [GET Cs] function key to obtain the
open/short circuit capacitance value (Cs).
TEST - M003 - EXAMPLE
MODE
01
GB
OSC
02
AC
03
04
DC
05
IR
PA
06
LC
07
Operation
www.valuetronics.com
Display current = (current read) - (offset current)
If Measured Value < Value set in ¶ 2.3.1.9:
Display current =
If Measured Value > Value set in ¶ 2.3.1.9:
Display current = (current read) - (offset current)
If Measured Value < Value set in ¶ 2.3.1.10:
Display current =
If Measured Value > Value set in ¶ 2.3.1.10:
Display current = (current read) - (offset current)
SOURCE
LIMIT
RES.
30.0A
100.0mΩ
Open
50%
1.500kV
0.500mA
2.400kV
0.500mA
1.000kV
10.0GΩ
CHANGE LEAD
0.300kV
10.00mA
GB
Remote
Lock
TEST - M003 - EXAMPLE
01
02
03
04
START
05
06
07
(
) (
2
−
current read
offset current
(
) (
−
2
current read
offset current
TEST - M003 - EXAMPLE
OFFSET
01
GET Cs
1.
PAGE UP
2.
PAGE DOW N
3.
PRESS START KEY TO GET Cs
SCANNER - 1
1 2 3 4 5 6 7 8
Offset
Error
MODE
SOURCE
LIMIT
RES.
OS
0.100kV
0.241nF
Pass
Remote
Lock
150799 Rev A3
)
2
)
2
MODE
SOURCE
LIMIT
RES.
GB
30.0A
100.0mΩ
MESSAGE
Please connect the DUT
.....
.....
Remote
OFFSET
GET Cs
PAGE UP
PAGE DOW N
SCANNER - 1
GB
1 2 3 4 5 6 7 8
Offset
Error
Page 111 of 177
OFFSET
GET Cs
PAGE UP
PAGE DOW N
SCANNER - 1
GB
1 2 3 4 5 6 7 8
Lock
Offset
Error