:IR
:LEVel <numeric value>
Command: :SOURce : SAFEty : STEP <n> : IR : LEVEL
This command sets the DC voltage level for the selected step.
<numeric value> is 50 to 1000 and the unit is volts.
:LEVel?
Command: :SOURce : SAFEty : STEP <n> : IR : LEVEL?
This command queries the AC voltage level for the selected step.
The return format is a numerical value: 50 ≤ value ≤ 1000 volts.
:LIMit
:TIME
Interface
www.valuetronics.com
:LOW <channel list>
Command: :SOURce : SAFEty : STEP <n> : DC : CHAN : LOW
(@Box#(<channel list>))
This command sets the low scan channels for the selected step.
:LOW?
Command: :SOURce : SAFEty : STEP <n> : DC : CHAN : LOW?
This command queries the low scan channels for the selected step.
Note:
Channel list must be 0 (zero) when no channels are set high or low.
:HIGH <numeric value>
Command: :SOURce : SAFEty : STEP <n> : IR : LIMit : HIGH
This command sets the high resistance limit for the selected step.
<numeric value> is 0 to 50G and the unit is ohms. 0 = OFF
:HIGH?
Command: :SOURce : SAFEty : STEP <n> : IR : LIMit : HIGH?
This command queries the high current limit for the selected step.
:LOW <numeric value>
Command: :SOURce : SAFEty : STEP <n> : IR : LIMit : LOW
This command sets the low resistance limit for the selected step.
<numeric value> is 0.1 to 50000M and the unit is ohms.
:LOW?
Command: :SOURce : SAFEty : STEP <n> : IR : LIMit : LOW?
This command queries the low current limit for the selected step.
:RAMP <numeric value>
Command: :SOURce : SAFEty : STEP <n> : IR : TIME : RAMP
This command sets the ramp time for the selected step.
<numeric value> is 0 to 999 and the unit is seconds. 0 = OFF
:RAMP?
Command: :SOURce : SAFEty : STEP <n> : IR : TIME : RAMP?
This command queries the ramp time for the selected step.
150799 Rev A3
Page 145 of 177
Need help?
Do you have a question about the Guardian 6100 Plus and is the answer not in the manual?