www.valuetronics.com
:CURRent
Command: :SOURce:SAFEty:STEP <n>:LC:POWer:CURRent
This command sets the power measurement to current for the selected step.
:LIMit[:HIGH] <numeric value>
Command: :SOURce:SAFEty:STEP<n>:LC:POWer:CURRent[:LIMit][:HIGH]<nv>
This command sets the high power limit for the selected step.
<numeric value> is 0 to 20 and the unit is amps
:LIMit[:HIGH[?
Command: :SOURce:SAFEty:STEP<n>:LC: POWer:CURRent[:LIMit][:HIGH]?
This command queries the high power limit for the selected step
The return format is a numerical value: 0 ≤ value ≤ 20A
:LIMit:LOW <numeric value>
Command: :SOURce:SAFEty:STEP<n>:LC: POWer:CURRent[:LIMit]:LOW <nv>
This command sets the low power limit for the selected step.
<numeric value> is 0 to 20 and the unit is amps.
:LIMit:LOW?
Command: :SOURce:SAFEty:STEP<n>:LC: POWer:CURRent[:LIMit]:LOW?
This command queries the low power limit for the selected step
The return format is a numerical value: 0 ≤ value ≤ 20A, 0 = OFF
:VA
Command: :SOURce:SAFEty:STEP <n>:LC:POWer:VA
This command sets the power measurement to volt-amps for the selected step.
:LIMit[:HIGH] <numeric value>
Command: :SOURce:SAFEty:STEP<n>:LC:POWer:VA[:LIMit][:HIGH]<nv>
This command sets the high power limit for the selected step.
<numeric value> is 0 to 4400 and the unit is volt-amperes
:LIMit[:HIGH]?
Command: :SOURce:SAFEty:STEP<n>:LC: POWer:VA[:LIMit][:HIGH]?
This command queries the high power limit for the selected step
The return format is a numerical value: 0 ≤ value ≤ 4400VA
:LIMit:LOW <numeric value>
Command: :SOURce:SAFEty:STEP<n>:LC: POWer:VA[:LIMit]:LOW <nv>
This command sets the low power limit for the selected step.
<numeric value> is 0 to 4400 and the unit is volt-amperes.
:LIMit:LOW?
Command: :SOURce:SAFEty:STEP<n>:LC: POWer:VA[:LIMit]:LOW?
This command queries the low power limit for the selected step
The return format is a numerical value: 0 ≤ value ≤ 4400VA, 0 = OFF