Offset Short Configuration
Figure 2-24 illustrates the short connection when the device under test is connected to the
G30 Corded Product Adapter. The Drive+ terminal is shorted to the ground of the
adapter.
Page 108 of 177
www.valuetronics.com
QuadTech
Guardian 6100 Plus
Electrical Safety Analyzer
1
0
NOTE:
Make the connection to the DRIVE+ and DRIVE-
terminals with the spade lug behind the nut.
Use the Bushing Driver Tool included to secure the
nut as shown here.
Figure 2-23: Offset SHORT Configuration
MENU
PRINT
1
4
7
0
CAL
UPDATE
DRIVE+ and DRIVE-
Output Terminals
SPADE lug
150799 Rev A3
TM
Twin
LOCAL
Por t s
DANGER
2
3
HIGH
VOLTAGE
CAUTION!
!
Max: 5kVAC
5
6
6kVDC
OUTPUT
PASS
FAIL
8
9
.
CLR
Drive
Sense Sense
Drive
+
-
ENTER
G15 Lead Set
SHORT
Operation