LC Current 10A Offset and Full
SAME connection of an AC current source (Freq = 60Hz) between the L-IN terminal and
the L-OUT terminal on the 6000-07 scanner.
Press [UP] to go to CAL step 9. (LCA 10A Offset
0.8A)
Press [STOP]
Press [START]
Using keypad enter the reading from the ammeter followed by [ENTER].
Press [STOP] to accept reading.
Press [UP] to go to CAL step 10. (LCA 10A Full
8A)
Press [STOP]
Press [START]
Using keypad enter the reading from the ammeter followed by [ENTER].
Press [STOP] to accept reading.
[LC BOARD] calibration is complete.
4.3.1.12
Finalize Calibration
When all calibration steps are complete:
Press [EXIT]
Release the [CAL ENABLE] switch to the OUT position using the tip of a small
screwdriver.
Cover [CAL ENABLE] switch with a calibration seal.
Service & Calibration
150799 Rev A3
Page 177 of 177