Language Reference
SOURce Subsystem
ESA E4405B: 0 dB to 56 dB in 8 dB steps
ESA E4407B, E4408B: 0 dB to 56 dB in 8 dB steps
Default Unit:
dB
Front Panel
Access:
Source, Attenuation Auto Man

Automatic Source Attenuation

:SOURce:POWer:ATTenuation:AUTO OFF|ON|0|1
:SOURce:POWer:ATTenuation:AUTO?
Selects if the source output level attenuator will be set automatically.
Factory Preset
and *RST:
On
Front Panel
Access:
Source, Attenuation Auto Man

Sets the Output Power

:SOURce:POWer[:LEVel][:IMMediate][:AMPLitude] <ampl>
:SOURce:POWer[:LEVel][:IMMediate][:AMPLitude]?
Specifies the source output power level. Use :SOURce:POWer:SWEep to
set the change in power level across the sweep. Also see
:SOURce:POWer:STARt and OUTPut[:STATe].
Factory Preset
−10 dBm
and *RST:
Range:
ESA E4401B, E4411B: –70 dBm to 3 dBm
ESA E4402B, E4403B: –66 dBm to 3 dBm
ESA E4404B: –66 dBm to 3 dBm
ESA E4405B: –66 dBm to 3 dBm
ESA E4407B, E4408B: –66 dBm to 3 dBm
Default Unit:
dBm
Front Panel
Access:
Source, Amplitude On Off

Sets the Source Output Power Mode

:SOURce:POWer:MODE FIXed|SWEep
:SOURce:POWer:MODE?