Omron FQ2-S Series User Manual page 70

Hide thumbs Also See for FQ2-S Series:
Table of Contents

Advertisement

Single Measurement
Example: Three Data Items Set for Parallel Data Output
Timing Chart
Run Mode entered.
ON
RUN signal
OFF
ON
TRIG signal
OFF
ON
BUSY signal
OFF
ON while measurements are being processed
OR signal
D0 to D15
signals
ON
GATE signal
OFF
1
The RUN signal turns ON when measurements are enabled and the Sensor is in Run Mode.
2
Turn ON the TRIG signal while the BUSY signal is OFF.
3
Measurement begins and the BUSY signal is turned ON during the measurement process.
4
When the measurement has been finished, the measurement results are output using an OR sig-
nal and the D0 to D15 signals, and the BUSY signal is turned OFF.
*1
You can also set the [BUSY output] parameter so that the BUSY signal is turned OFF after the completion of data logging, image logging,
or displaying results.
5
After the BUSY signal turns OFF, the GATE signal is turned ON when the time that is set in the
[GATE ON delay] parameter in the communications settings has elapsed.
6
The GATE signal is turned ON, and then the GATE signal is turned OFF when the time that is set
in the [Output time] parameter in the communications settings has elapsed.
*2
Set the GATE ON delay and output time for the GATE signal so that the total time does not exceed the output period.
Important
Data Output Time and TRIG Signal Input Interval
Set the input interval for the TRIG signal so that it is equal to or greater than the total output time. If the input interval
for the TRIG signal is shorter than the total output time, the output data buffer will eventually overflow and output data
will be discarded.
68
Controlling Operation and Outputting Data with a Parallel Interface Sensor Data Unit
ON for 1 ms min.
(depends on BUSY output conditions)
Turned ON when overall judgement is NG.
(Polarity of all output signals: Positive)
Data 0
GATE ON delay
Output time
Output period
The total output time is as follows: Output period × Number of output data items.
Setup Mode entered.
Data 1
Data 2
*1
*2
*2
FQ2-S/CH User's Manual
for Communications Settings