18
nanoCell
Measurements
286
Evolution 200 Series User Guide
Note Install the lid immediately after loading the sample to prevent evaporation of the
sample.
Figure 6.
The nanoCell accessory with the lid installed.
Related Topics
Installing the nanoCell
Measurements
Cleaning the nanoCell
Storing the nanoCell
Pathlength Correction Factors
Frequently Asked Questions
Always collect the zero or baseline measurement with a blank solution in the nanoCell.
Note If a high background signal or inconsistent data is observed, remeasure the blank
solution for the baseline or zero. Some measurement errors are due to a poor baseline or
zero.
Related Topics
Installing the nanoCell
Loading a sample
Cleaning the nanoCell
Storing the nanoCell
Pathlength Correction Factors
Frequently Asked Questions
Thermo Scientific