28
Smart Sipper
412
Evolution 200 Series User Guide
To install the Smart Sipper
1. Open the sample compartment. See
Compartment
Doors.
2. Remove the sample compartment front panel. See
3. Remove any existing cell holder or accessory.
4. Install the accessory. See
The software displays a prompt to initialize the accessory.
CAUTION Avoid pinch hazard. Keep hands and objects clear of the accessory during
initialization.
Languages
ATTENTION Évitez tout risque de pincement. Gardez vos mains, ainsi que tout objet,
à l'écart de l'accessoire pendant l'initialisation.
VORSICHT Quetschungen sind zu vermeiden. Hände und Objekte sind während der
Initialisierung vom Zubehör fern zu halten.
CAUTELA Evitare il rischio di pizzicamento. Tenere mani e oggetti lontano
dall'accessorio durante l'inizializzazione.
PRECAUCIÓN Evite riesgos de atrapamiento. Mantenga manos y objetos lejos del
alcance del accesorio durante la inicialización.
5. Click OK to initialize the accessory.
Initialization reads information about the accessory and reset the pump motors.
6. Install the Sipper sample compartment front panel.
Align the tab on the bottom edge with the slot in the instrument cover and press the
panel in place.
Opening and Closing the Sample and Reference
Installing a Smart