Water Level Failure
Thermo Scientific
If an alarm occurs
1. Determine the cause of the alarm by consulting the recirculator front panel display.
2. Press the right (Reset) button, if available, on the recirculator front panel.
See Manual reset prealarm in the
display in Alarm Mode.
3. Turn off the power switch on the recirculator back panel.
4. Fix the problem. See
5. Turn on the recirculator power switch.
6. If the alarm occurs again, turn off the power switch and contact technical support. See
Contacting
Us.
Related Topics
Water Level Failure
Other System Failures
Preinstallation
Installation
Operation
Maintenance
The accessory is equipped with a water level sensor that detects small leaks.
If a water level failure occurs
1. Press the Reset button, if available, on the recirculator front panel.
See Manual reset prealarm in the
display in Alarm Mode.
2. Turn off the power switch on the recirculator back panel.
3. Open the sample compartment and clean up any liquids that may have leaked from the
accessory.
4. Tighten all connectors to and from the recirculator.
5. Fill the reservoir. See
6. Turn on the recirculator power switch.
Parameters Menu
Water Level Failure
and
Other System
Parameters Menu
Filling the
Reservoir.
TPS-1500W Sealed Peltier Recirculator
to configure the Reset option to
Failures.
to configure the Reset option to
Evolution 200 Series User Guide