How the Type Calculates Sample Concentration
Manually entered
Uses the equation C = K * A, where C is the concentration of the
factor
analyte in the selected units, K is the entered factor, and A is the
absorbance.
Measure single
Divides the sample absorbance value by the average absorbance value
standard
of a single standard. This is in effect a "Standard curve" quantification
(see below) with just one standard.
Standard curve
Uses a simple standard curve based on the absorbance of a set of
standards at a specified wavelength.
Standard curve with
Takes the average of values determined by each of two standard curves
two wavelengths
associated with the specified wavelengths.
Advanced standard
Uses a standard curve generated from a specified equation.
curve
Advanced without
Uses a "single result equation" specified on the
standards
(Calculates the concentration or answer directly from this equation.)
Related Topics
Quant
Measurement Tab for Quant
Standards Tab for Quant
Instrument Tab for Quant
Samples Tab for Quant
Accessories Tab for Quant
Formulas and Units Tab
Quantifying Sample Data
Performing a Quantitative Analysis
Measurement Tab for Quant
Click Settings in
Quant
in the right pane. The available features depend on the
Quant Type setting on the
each type are described in tables below.
2
INSIGHT Software
Measurement
tab.
to display the Measurement tab
Type
tab. The features for
Evolution 200 Series User Guide
Operation
53