20
Peltier Thermostatted Single Cell Holder
320
Evolution 200 Series User Guide
Cable
6. Install the cell holder in the spectrophotometer sample compartment. See
Manual
Accessory.
7. Feed the other end of the cable through the opening behind the tubing access panel. See
Tubing Access
Panel.
8. Connect the free end of the cable to the Cell Holder socket on the back of the
temperature controller.
9. Use appropriate tubing and the provided adapters to connect the two fluid ports on the
cell holder to a water pump or fluid recirculator. See
Accessory.
Ensure the tubing does not obstruct the light beam.
Cell Holder