Thermo Scientific Evolution 200 Series User Manual page 433

Uv-vis instruments
Table of Contents

Advertisement

Thermo Scientific
Feature
Sipper Workflows
Manual Control
Sip Volume button
Pump On/Pump Off buttons
Related Topics
Safety Considerations
Operating Precautions
Installation
Calibration
Maintenance
Description
Defines automated sipper workflows. Use the Measure
button in the software to start the workflow.
• Sip, settle, measure, prompt, return draws the
specified sip volume, delays the specified settle time,
measures the sample, and prompts to move the inlet
tube to the return container. When the user clicks OK
or presses the Sipper front panel switch, returns the
sample to the container.
• Sip, draw air, settle, measure, prompt, rinse, draw
air draws the specified sip volume and prompts to
remove the inlet tube. When the user clicks OK or
presses the front panel switch, draws air for the
specified period, waits the specified settle time,
measures the sample and prompts to place the inlet
tube in the rinse solution. When the user clicks OK or
presses the switch, draws the rinse solution for the
specified time and prompts to remove the inlet tube.
When the user clicks OK or presses the switch, draws
air for the specified period.
Samples are measured using the selected application.
Results are placed in the selected workbook.
Activates the Sipper front panel switch as a pump control.
When the check box is selected, pressing the switch once
causes the pump to draw through the inlet tube; pressing it
again stops it.
Immediately causes the pump to draw the specified sip
volume through the inlet tube.
Toggles between Pump On and Pump Off. Pump On
causes the pump to draw continuously through the inlet
tube; Pump Off causes it to stop.
Smart Sipper
Evolution 200 Series User Guide
28
419