Replacing Pump Tubing
Cleaning Flow Cells
Thermo Scientific
To clean the accessory exterior
1. Remove the accessory from the sample compartment. See
2. Use a lint-free cloth dampened with plain water to wipe the area near the cell holder.
3. Dry the surface with another cloth.
Related Topics
Routine Maintenance
Replacing Pump Tubing
Cleaning Flow Cells
Replacing Flow Cell Tubing
Maintenance
Examine pump tubing weekly and when pumping efficiency is impaired. Replace crimped or
worn tubing (see
Installing Pump
If you install a different type or size tubing, recalibrate before analyzing samples (see
Calibration).
Related Topics
Routine Maintenance
Cleaning the Accessory
Cleaning Flow Cells
Replacing Flow Cell Tubing
Maintenance
Rinse the flow cell and inlet tubing thoroughly with laboratory grade purified water after each
analysis and before turning off the Smart Sipper. Keep the cell filled with purified water when
not in use.
At least once a week, remove the flow cell from the accessory and soak it in a solution of
Decon™ detergent (or equivalent), diluted according to the manufacturer's instructions. When
finished, reconnect the cell and flush it thoroughly with laboratory grade purified water.
Tubing).
Smart Sipper
Installing a Smart
Accessory.
Evolution 200 Series User Guide