Measurement Tab for Pierce Modified Lowry
Thermo Scientific
Feature
Standard curve
Pathlength
Related Topics
Measurement Tab for Pierce Modified Lowry
Standards Tab for Pierce Modified Lowry
Instrument Tab for Pierce Modified Lowry
Accessories Tab for Pierce Modified Lowry
Samples Tab for Pierce Modified Lowry
Feature
Description
Analysis wavelength Wavelength(s) to use for the analysis.
Correction
Specifies bichromatic normalization of the absorbance data.
To use uncorrected data, select None. Without baseline correction,
spectra may be offset from the baseline. If this offset is significant, the
calculated protein concentration may be higher than the true value.
Single point subtracts the sample absorbance at the entered
wavelength from the absorbance at every data point.
Sloping baseline draws a line between the data points at the entered
wavelengths, finds its absorbance value at the analytical wavelength,
and subtracts it from the measured absorbance.
Component name
Component to quantify.
Description
Type of quantitative analysis used.
Distance the light travels through the sample.
Click Settings in Protein Lowry to display the
Measurement tab in the right pane.
4
Life Science Applications
Pierce Modified Lowry
Evolution 200 Series User Guide
203