Evolution 200 Series User Guide
To install the nanoCell, insert the accessory into a single cell holder in the sample light path of
the spectrophotometer. One of the quartz windows on the side of the nanoCell accessory (see
Figure
2) should face the light beam. The nanoCell accessory can be installed with either side
of the nanoCell facing the light source.
Figure 2.
The quartz windows of the nanoCell allow light to enter and exit this accessory. Align
these windows in the path of the beam.
Quartz window
Tip To ensure the highest precision data, do not remove the nanoCell from the cell holder
until all measurements are complete
NOTICE Install the nanoCell only in accessories designed to hold a single cell. Using the
nanoCell in a multi-cell holder could lead to inaccurate data.
Related Topics
Loading a sample
Measurements
Cleaning the nanoCell
Pathlength Correction Factors
Frequently Asked Questions
To load a sample on the nanoCell, use a small volume pipette to place between 0.70 μl and
3.0 μl of solution on the sample loading window.