Thermo Scientific Evolution 200 Series User Manual page 425

Uv-vis instruments
Table of Contents

Advertisement

Installing the Smart Sipper
Thermo Scientific
4. Press the tubing into both clamps.
NOTICE Leave at least 2 cm of tubing extending beyond each clamp.
Rotate pump head
Open clamps
5. Insert a nipple (provided) into each end of the tubing.
NOTICE Remove the tubing from the pump before storing the accessory. Leave the
instrument and accessory powered on when the pump is idle.
Related Topics
Installing the Smart Sipper
Installing a Flow Cell
Installation
It is not necessary to power off the instrument when installing or removing a Smart Accessory.
Insert tubing
Nipple
Smart Sipper
Pump In
Pump Out
Evolution 200 Series User Guide
28
411