Feature
Record temperature with mea-
surement
Use temperature controller
Start measurements at target
temperature
Match probes to cell changer
positions
Related Topics
Configuring a Target Temperature
Displaying the Temperature Status Monitor
Using INSIGHT Software to Set Temperature and Stir Speed
Configuring a Target Temperature
To configure a target temperature
1. Select Start measurement at target temperature.
2. Enter the Target temperature and Tolerance in °C.
• For a temperature probe.
To start the measurement when the probe reaches the target temperature, set Start on
to Probe temperature and then select the Probe channel. The number of channels
in the list box will match the available ports on the temperature probe interface box.
The probe channel, current temperature and target range appear during the
measurement.
Description
If a temperature probe is used, records the temperature of
the sample after each measurement. Temperature readings
are stored in the sample properties.
If a Peltier cell holder or changer is used, defines a setpoint
temperature for the cell block and a stir speed if desired.
The available settings depend on the specifications for the
temperature-controlled accessory.
Starts the measurement when the sample (requires a tem-
perature probe) or cell block (requires a Peltier cell holder
or changer) reaches a specified temperature.
If a temperature probe is used, maps probe channels 1
through 8 with up to eight cell changer positions. Useful
only when using an automatic cell changer with multiple
temperature probes.
Deselect this to always record the temperature of probe 1.
SPE-8W Smart Peltier Linear Cell Changer
Evolution 200 Series User Guide
26