Manually entered factor
Feature
Analysis wavelength(s)
Correction
Component name
Unit
Use user-defined factor
Use extinction coefficient Calculates component concentration with absorbance (A) equal
Description
Wavelength(s) to use for the quantitative analysis.
To use uncorrected data, select None.
Single point subtracts the sample absorbance at the entered
wavelength from the absorbance at every data point.
Sloping baseline draws a line between the data points at the
entered wavelengths, finds its Y value at the analytical
wavelength, and subtracts it from the measured absorbance.
Component to quantify.
Component concentration unit. The available units depend on
the
Formulas & Units tab
Calculates component concentration using the entered factor (K
in the equation C = K *A), with the selected unit.
to
ε
l c
where ε is the entered extinction coefficient (a constant
dependent on the component and wavelength), c is
concentration in the selected unit, and
pathlength).
settings in Options.
is 1 cm (the