Thermo Scientific Evolution 200 Series User Manual page 148

Uv-vis instruments
Table of Contents

Advertisement

3
Evolution 200 Series Spectrophotometers
Maintenance
134
Evolution 200 Series User Guide
.
Thumbscrew
4. Replace the detector access cover.
5. Store the detector module in a secure, dust-free location.
To install the detector module
1. Turn off the spectrophotometer. See
2. Remove the detector access cover. See
3. Unplug the accessory detector cable if necessary.
4. Position the detector module so its thumbscrew faces left.
Thumbscrew
5. Line up the connector on the bottom of the detector module with the connector in the
detector compartment, and then press down on the module to secure the connection.
6. Tighten the thumbscrew by hand.
7. Replace the detector access cover.
Related Topics
Routine Maintenance
Cleaning the Instrument
Cleaning the Touchscreen
Using the
Spectrophotometer.
Detector Access
Cover.
Detector module
Thermo Scientific