9. Set Number of samples to the number of samples to be measured.
10. Use Load Samples to import the sample list or enter sample names and descriptions if
desired. See Settings > Samples to create the list ahead of time.
11. Click OK.
The loading guide shows the measurement sequence for the cell changer as defined in the
sample list.
12. Place the cuvettes in the correct cell changer positions.
Note If the application requires a blank and it was not measured in step 4, the first
cell changer position will be measured as the blank.
13. Click OK or press Run on the instrument keypad.
To cancel, click Cancel or press Stop on the instrument keypad.
If a blank is required, it will be measured first, followed by the samples in order.
If the sample list has more entries than there are spaces in the cell changer, the Loading
Guide reappears after the last cuvette is measured.
14. If the loading guide appears, remove all cuvettes from the cell changer and fill it with the
next set of samples as defined in the loading guide.
15. Click OK.
Smart Rotary Cell Changer
Evolution 200 Series User Guide
27