Acoustic Alarm - Thermo Scientific Evolution 200 Series User Manual

Uv-vis instruments
Table of Contents

Advertisement

19
TPS-1500W Sealed Peltier Recirculator
Troubleshooting

Acoustic Alarm

308
Evolution 200 Series User Guide
3. Turn off the power switch on the recirculator back panel.
4. Unplug any tubing from the inlet connector on the recirculator back panel and connect
the tube from the fill bottle.
5. Turn the fill bottle upside down and squeeze it until liquid escapes through the vent.
The reservoir holds approximately 150 mls.
6. Disconnect the fill bottle and reconnect the tubing.
7. Turn on the recirculator power switch.
8. If a water level alarm occurs, turn off the recirculator and repeat steps 3 through 7.
9. If the water level is okay, open the vent to evacuate any air bubbles in the lines.
10. Close the vent.
Leave the vent closed while the recirculator is in use.
Related Topics
Routine Maintenance
Cleaning the Accessory
Preinstallation
Installation
Operation
Troubleshooting
System alarms indicate failure of the accessory due to malfunction or incorrect use. Alarms
may be visible on the display, indicated acoustically or communicated via RS-232 or USB
connections.
Contents
Acoustic Alarm
Water Level Failure
Other System Failures
The accessory sends an acoustic alarm when it detects a leak or if another failure occurs.