Thermo Scientific Evolution 200 Series User Manual page 452

Uv-vis instruments
Table of Contents

Advertisement

29
Temperature Probe
Routine Maintenance
Cleaning Temperature Probes
Replacing Stopper O-Rings
438
Evolution 200 Series User Guide
• Clean probes immediately after each measurement and before storing them. See
Temperature
Probes.
• Check stopper O-rings regularly for wear and replace them when necessary. See
Stopper
O-Rings.
Note Use of the temperature probe with corrosive samples or extreme temperatures can
cause O-rings to break down faster.
Related Topics
Cleaning Temperature Probes
Replacing Stopper O-Rings
Maintenance
Clean temperature probes frequently to avoid contaminating subsequent samples. Wipe off
spilled chemicals immediately.
To clean a probe, rinse the tip with water and then with methanol or isopropyl alcohol and air
dry.
Related Topics
Routine Maintenance
Replacing Stopper O-Rings
Maintenance
Check stopper gaskets and O-rings regularly for wear and replace them when necessary. The
square stopper has two gaskets and internal O-ring. The round stopper has an internal
O-ring.
A 4 mm screw (at least 3 mm long) is required to remove the liner that supports the internal
O-ring.
To replace the internal O-ring (round or square stopper)
1. Insert the 4 mm screw several turns into the stopper base, and use it to pull the liner out
of the base (leave the screw in the liner).
Cleaning
Replacing