18
nanoCell
290
Evolution 200 Series User Guide
Q5
What needs to be cleaned after each use? The top of the nanoCell only, or the inner
part of the lid as well?
A5
To eliminate sample carry-over or mixing two liquids together and improve accuracy,
clean the inner part of the lid between measurements. See
Q6
Can I use solvents to clean the surface of the nanoCell? If so, which solvents?
A6
The most suitable solvents for cleaning are methanol, ethanol, and isopropanol. Do
not use strong acids and bases or any of the following solvents with the nanoCell:
ethylamine, diethylene triamine, methylene chloride, and phosphoric acid. See
Cleaning the
nanoCell.
Related Topics
Using the nanoCell
Cleaning the nanoCell
Storing the nanoCell
Pathlength Correction Factors
Cleaning the
nanoCell.