SPE-8W Smart Peltier Linear Cell Changer
378
Evolution 200 Series User Guide
CAUTELA Evitare il rischio di pizzicamento. Tenere mani e oggetti lontano
dall'accessorio durante l'inizializzazione.
PRECAUCIÓN Evite riesgos de atrapamiento. Mantenga manos y objetos lejos del
alcance del accesorio durante la inicialización.
7. Feed the other end of the cable through the opening behind the tubing access panel. See
Tubing Access
Panel.
8. Connect the free end of the cable to the Cell Holder socket on the back of the
temperature controller.
9. Use appropriate tubing and the provided adapters to connect the two fluid ports on the
cell holder to a water pump or fluid recirculator. See
Accessory.
Ensure the tubing and cable do not obstruct the light beam. Use the provided clips if
necessary and test all cell changer positions.
10. Replace the sample compartment front panel. See
11. Close the sample compartment cover.
12. Turn on the power switch on the controller back panel.
Installing a Liquid Thermostatted
Water ports
Sample Compartment Front
Panel.
Thermo Scientific