Installing a Smart Accessory

Thermo Scientific
2. Remove any accessory from the compartment (see the instructions below).
3. Position the accessory so the two alignment holes on its baseplate fit over pins A and B in
the sample compartment.
4. Gently push down on the front edge of the baseplate until it snaps in place under the
toggle.
5. Close the sample compartment door. See
Compartment
Doors.
For operating instructions and installation details, see the accessory name under Accessories.
To remove a manual accessory
1. Use the handle or finger hole on the accessory baseplate to release the plate from the
toggle.
2. Lift the plate off the alignment pins and remove the accessory from the sample
compartment.
Related Topics
Installing a Liquid Thermostatted Accessory
Installing a Smart Accessory
Installing a Reference Cell Holder
Smart Accessories include cell changers and sample holders that feature auto recognition,
smart alignment and serial number reporting. Smart Accessories ensure software methods are
properly configured.
These accessories install in the sample compartment. A connector under the accessory
baseplate provides data communication and power.
It is not necessary to power off the instrument while installing or removing a Smart Accessory.
3
Evolution 200 Series Spectrophotometers
Opening and Closing the Sample and Reference
Evolution 200 Series User Guide
Accessories
121