Samples Tab for Quant
Feature
Description
Number of samples Number of samples in the analysis.
Base name
Beginning of the name for sample spectra, to which a sequential
number is appended. For example, the names of three sample spectra
using the base name "dye" would be dye1, dye2 and dye3.
Sample averaging
Whether and how to average concentration values from multiple
measurements of the same sample or from multiple samples. To
average multiple measurements of the same sample, select Replicate
and set Number of replicates to the total number of measurements.
To average values from two or three samples, select Duplicate or
Triplicate, respectively.
Use concentration
Displays high and low limit lines on the Run Chart tab to show
limits
whether sample concentrations are within the specified limits.
Use weight/volume
If available for the current unit, corrects sample concentrations using
correction
the entered target weight and volume:
corrected concentration = measured concentration * (actual weight /
target weight) * (target volume / actual volume)
Related Topics
Quant
Type Tab for Quant
Measurement Tab for Quant
Standards Tab for Quant
Instrument Tab for Quant
Accessories Tab for Quant
Formulas and Units Tab
Quantifying Sample Data
Performing a Quantitative Analysis
Click Settings in
Quant
the right pane.
2
INSIGHT Software
to display the Samples tab in
Evolution 200 Series User Guide
Operation
67