4
Life Science Applications
Nucleic Acid Labels
Instrument Tab for Nucleic Acid Labels
154
Evolution 200 Series User Guide
Accessories Tab for Nucleic Acid Labels
Samples Tab for Nucleic Acid Labels
Feature
Mode
Data mode
Start wavelength and End
wavelength (Scan only)
Bandwidth
Integration time
Data interval (Scan only) Difference in wavelength between two consecutive data points.
Scan speed (Scan only)
Estimated time (Scan
only)
Table of wavelengths
(Fixed only)
Related Topics
Type Tab for Nucleic Acid Labels
Measurement Tab for Nucleic Acid Labels
Accessories Tab for Nucleic Acid Labels
Samples Tab for Nucleic Acid Labels
Click Settings in
Instrument tab in the right pane.
Description
Specifies Scan or Fixed data acquisition.
Y-axis format for acquired data.
Starting and ending values of the wavelength range for acquiring
data.
A larger spectral bandwidth increases the amount of energy that
passes through the sample. Increasing the bandwidth can
improve the signal-to-noise ratio; however, a larger spectral
bandwidth decreases the ability to resolve closely spaced spectral
features.
How long the system acquires and averages data at each data
interval (for scanning measurements) or at each measured
wavelength (for fixed-wavelength measurements). Increasing the
integration time improves the signal-to-noise ratio but reduces
the scan speed.
Wavelength range covered per unit time. Varies inversely with
integration time. Increases as the data interval increases.
Estimated duration of data acquisition. Increases as integration
time increases and as the data interval decreases. Varies inversely
with scan speed.
Shows the analysis wavelength(s) entered on the Measurement
tab.
Nucleic Acid Labels
to display the
Thermo Scientific