SPE-8W Smart Peltier Linear Cell Changer
384
Evolution 200 Series User Guide
• For a temperature-controlled accessory.
To start the measurement when the cell block reaches the target temperature, set
Start on to Block temperature.
Only the current temperature and target range appear during the measurement.
The Start button is disabled since the measurement is configured to start automatically at
the target temperature.
3. To allow time for equilibration, select Wait for equilibration and enter the duration.
The remaining equilibration time is displayed during the measurement.
If the temperature moves outside the target range, the countdown resets when the
temperature is again within range.
The Start button becomes active when the temperature remains in the target range for the
entire equilibration period.
4. To start the measurement automatically after equilibration, select Start measurement
after equilibration. This disables the Start button in the status dialog box. The
measurement starts automatically when the temperature remains in the target range
during the entire equilibration period.
Related Topics
Temperature and Stirrer Settings