6. Click Next.
A prompt appears. The contents depends on the Prompting setting.
7. Record the volume of liquid in the cylinder.
8. Click OK or Pump Now (or wait until the counter is finished).
The Sipper attempts to draw the specified volume, and then a prompt requests the actual
volume drawn.
9. Remove the inlet tube from the cylinder and record the volume of liquid remaining.
10. Subtract the volume remaining in the cylinder from the initial volume to calculate the
actual volume drawn through the inlet tube. Enter this value in the Actual box and click
Next.
11. Repeat steps 7 through 10 until the Actual volume matches the Expected volume.
12. Click Done.