Thermo Scientific
Feature
Description
Integration time
How long the system acquires and averages data at each data interval.
Increasing the integration time improves the signal-to-noise ratio but
reduces the scan speed. For scan experiments only.
Data interval
Difference in wavelength between two consecutive data points. For
scan experiments only.
Scan speed
Wavelength range covered per unit time. Varies inversely with
integration time. Increases as the data interval increases. For scan
experiments only.
Estimated time
Estimated duration of data acquisition. Increases as integration time
increases and as the data interval decreases. Varies inversely with scan
speed. For scan experiments only.
Wavelength table
Wavelengths to monitor during the reaction. For single and multiple
wavelength experiments only.
Related Topics
Type Tab for Rate
Measurement Tab for Rate
Accessories Tab for Rate
Samples Tab for Rate
Rate Calculations Tab for Rate
Performing a Kinetics Measurement
Accessories Tab for Rate
The status of accessories can be monitored
Related Topics
Accessories
Instrument Status Monitors
Type Tab for Rate
Measurement Tab for Rate
Click Settings in
Rate
to display the Accessories tab in
the right pane. The available parameters depend on the
installed accessories. See
during measurements.
2
INSIGHT Software
Accessories
for details.
Evolution 200 Series User Guide
Operation
71