SPE-8W Smart Peltier Linear Cell Changer

Preinstallation

Verifying the Power Supply and Fuse

374
Evolution 200 Series User Guide
Related Topics
System Connections
External Connections
Temperature Controller Front and Rear Panel
Review this section before installing the accessory.
Contents
Verifying the Power Supply and Fuse

Changing the Voltage Setting or Fuse

Related Topics
General Safety Information
Product Description
Installation
Operation
Maintenance
Troubleshooting
NOTICE The unit is factory set to operate at 220 V.
For operation at 120 V, configure the voltage selector on the Peltier back panel and insert the
correct fuse as described in the following table.
Power Supply
Type of fuse - 250 V
240 V
T 1A, Slo Blo
120 V
T 2A, Slo Blo
Adjustment for frequency change between 50 and 60 Hz is not necessary.