We provide two kinds of pump tubing with the Smart Sipper: silicone rubber (Standard
Tubing Pack) and Viton® (Acid Tubing Pack).
Verify tubing size and sample compatibility prior to installation. See Specifications.
CAUTION Avoid pinch hazard. Do not install pump tubing while the instrument is
powered on.
Languages
ATTENTION Évitez tout risque de pincement. N'installez pas la tubulure de la pompe
lorsque l'instrument est sous tension.
VORSICHT Quetschungen sind zu vermeiden. Die Pumpenschläuche dürfen nicht
installiert werden, während das Instrument eingeschaltet ist.
CAUTELA Evitare il rischio di pizzicamento. Non installare le tubazioni della pompa
mentre lo strumento è acceso.
PRECAUCIÓN Evite riesgos de atrapamiento. No instale los tubos de la bomba con el
instrumento encendido.
To install pump tubing
1. Remove the Smart Sipper from the spectrophotometer or turn off the spectrophotometer
main power switch.
2. Cut an 18 cm length of pump tubing.
3. Taking care not to stretch, slacken or twist the tubing, feed it counterclockwise around
the track while manually rotating the pump head.