Calibrated Validation Carousel (Cvc) - Thermo Scientific Evolution 200 Series User Manual

Uv-vis instruments
Table of Contents

Advertisement

Calibrated Validation Carousel (CVC)

Test
Wavelength Accuracy
Wavelength Repeatability
Absorbance Accuracy
UV Absorbance Accuracy
Absorbance Repeatability
Stray Light
Noise
Thermo Scientific
The Calibrated Validation Carousel (CVC) is an automated performance verification
accessory.
NOTICE
• Do not touch the exposed surface of the standards in the CVC.
• Do not allow moisture to leak into the accessory.
Contents
• Installing the Carousel on the Smart Base Unit
• Installing Complete CVC in the Instrument
• CVC Test Limits
• Using the CVC
Maintenance
The CVC carousel uses the same Smart base unit as the Smart Thermostatted Rotary 7-Cell
Changer (Rotary Cell Changer). The CVC is also supplied without the base unit for
customers wishing to use the CVC with an existing Rotary Cell Changer base.
The CVC accurately positions the required standards for each test in the sample beam. The
following table describes the tests and test materials included with the CVC.
Standard
Holmium perchlorate
Holmium perchlorate
1A Neutral Density filter
Potassium dichromate
1A Neutral Density filter
220, 340 nm cut-off filters Sodium Iodide
0A and 1A Neutral Density
filter
INSIGHT software stores all instrument performance specifications and tolerances.
USP
Holmium perchlorate
Holmium perchlorate
Potassium dichromate
Potassium dichromate
Potassium dichromate
Instrument lamp
Evolution 200 Series User Guide
12
EP
Holmium perchlorate
Holmium perchlorate
Potassium dichromate
Potassium dichromate
Potassium dichromate
Potassium chloride
Instrument lamp
249

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Evolution 201Evolution 220Evolution 260 bio

Table of Contents