Thermo Scientific Evolution 200 Series User Manual page 84

Uv-vis instruments
Table of Contents

Advertisement

2
INSIGHT Software
Operation
70
Evolution 200 Series User Guide
Feature
Description
Number of cycles
Number of measurement cycles. For scan experiments only.
Duration
Length of the experiment, calculated by multiplying the cycle time
by the number of cycles (see above). For scan experiments only.
Related Topics
Type Tab for Rate
Instrument Tab for Rate
Accessories Tab for Rate
Samples Tab for Rate
Rate Calculations Tab for Rate
Performing a Kinetics Measurement
Instrument Tab for Rate
Feature
Description
Data mode
Y-axis format for acquired data.
For Abs*Factor, set Factor to a value by which to multiply measured
absorbance values. For scan experiments only.
Smooth
Select the desired degree of smoothing. Smoothing reduces noise in
the data but can also remove some spectral features such as small
peaks, valleys and shoulders. For more information, see
Smoothing. For scan experiments only.
Derivative
Select the desired derivative order for acquired data. For scan
experiments only.
Start wavelength
Starting and ending values of the wavelength range for acquiring
and End wavelength
data. For scan experiments only.
Bandwidth
A larger spectral bandwidth increases the amount of energy that
passes through the sample. Increasing the bandwidth can improve the
signal-to-noise ratio; however, a larger spectral bandwidth decreases
the ability to resolve closely spaced spectral features, especially in
scanning measurements.
Click Settings in
Rate
to display the Instrument tab in
the right pane.
About
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Evolution 201Evolution 220Evolution 260 bio

Table of Contents