Thermo Scientific Evolution 200 Series User Manual page 206

Uv-vis instruments
Table of Contents

Advertisement

4
Life Science Applications
Protein Bradford
Type Tab for Protein Bradford
Measurement Tab for Protein Bradford
192
Evolution 200 Series User Guide
Feature
Name
Description (optional) Description of the template.
Standard curve
Pathlength
Related Topics
Measurement Tab for Protein Bradford
Standards Tab for Protein Bradford
Instrument Tab for Protein Bradford
Accessories Tab for Protein Bradford
Samples Tab for Protein Bradford
Feature
Description
Analysis wavelength Wavelength(s) to use for the analysis.
Correction
Specifies bichromatic normalization of the absorbance data.
To use uncorrected data, select None. Without baseline correction,
spectra may be offset from the baseline. If this offset is significant, the
calculated protein concentration may be higher than the true value.
Single point subtracts the sample absorbance at the entered
wavelength from the absorbance at every data point.
Sloping baseline draws a line between the data points at the entered
wavelengths, finds its absorbance value at the analytical wavelength,
and subtracts it from the measured absorbance.
Component name
Component to quantify.
Click Settings in
tab in the right pane.
Description
Application name.
Type of quantitative analysis used.
Distance the light travels through the sample.
Click Settings in
Measurement tab in the right pane.
Protein Bradford
to display the Type
Protein Bradford
to display the
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Evolution 201Evolution 220Evolution 260 bio

Table of Contents