System Settings; Alignment Tab - Thermo Scientific Evolution 200 Series User Manual

Uv-vis instruments
Table of Contents

Advertisement

System Settings

Alignment Tab

Thermo Scientific
Alignment Tab
Calibrations Tab
Lamp Tab
System Tab
Diagnostics tab
Note The Diagnostics tab is for our use only.
Use the Alignment tab in
while aligning an accessory.
To perform an alignment
1. To measure intensities using a green light (510 nm), select Green. To use no filter, select
White.
The correct setting depends on the installed accessory. See Accessories.
2. Set Bandwidth as desired.
The available settings depend on the instrument.
3. Click Start.
Every half second the sample and reference detector intensities are plotted and displayed
by the Sample beam and Reference readouts.
These are available in System Settings:
System Settings
to view sample and reference detector intensities
2
INSIGHT Software
System Settings
Evolution 200 Series User Guide
99