Calibrated Validation Carousel (CVC)
CVC Test Results
Viewing CVC Test Reports
256
Evolution 200 Series User Guide
Maintenance
For each test in the suite, the report shows:
• Test name
• Operator
• Time and date of the test
• Instrument type
• Instrument serial number
• CVC serial number
• Expected result or range, measured result and a pass or fail result
Related Topics
Installing the Carousel on the Smart Base Unit
Installing Complete CVC in the Instrument
Loading CVC Calibration Data
CVC Test Limits
Running Performance Verification Tests
Viewing CVC Test Reports
Maintenance
To view a CVC test report
1. From the INSIGHT software Home screen, click My Data.
2. Click the PV Reports tab.
The right pane shows all the performance tests run with the instrument.