Water Level Failure; Other System Failures - Thermo Scientific Evolution 200 Series User Manual

Uv-vis instruments
Table of Contents

Advertisement

19
TPS-1500W Sealed Peltier Recirculator

Other System Failures

310
Evolution 200 Series User Guide
7. If the water level alarm occurs again, turn off the power switch, fill the reservoir again and
turn on the power switch.
Note You may need to fill the reservoir several times before the water level alarm resets.
Related Topics
Acoustic Alarm
Other System Failures
Preinstallation
Installation
Operation
Maintenance
Problem
Possible Cause
Temperature or
Recirculator
memory failure
malfunction or
incorrect use
Related Topics
Acoustic Alarm

Water Level Failure

Preinstallation
Installation
Operation
Maintenance
Solution
Immediately turn off the main power switch
and then turn it on again. If the failure
reoccurs, turn off the power and contact
technical support. See
Contacting
Us.