[:Sense]:Sweep Subsection - Agilent Technologies ESA-E Series Programmer's Manual

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Front Panel
Access:
Input/Output (or Input), Input Mixer, Signal Ident On Off

[:SENSe]:SWEep Subsection

Sweep Time
[:SENSe]:SWEep:TIME <time>
[:SENSe]:SWEep:TIME?
Specifies the time in which the instrument sweeps the display.
Factory Preset
and *RST:
ESA E4401B, E4411B: 4ms
ESA E4402B, E4403B: 5 ms
ESA E4404B: 16.75 ms
ESA E4405B: 33 ms
ESA E4407B, E4408B: 265 ms
Range:
The range depends upon the installed options, number
of sweep points, and firmware revision of your
instrument. See "Sweep Time Range" in the
Specifications Guide for details.
Default Unit:
seconds
Remarks:
A span value of 0 Hz causes the analyzer to enter zero
span mode. In zero span the X-axis represents time
rather than frequency. In this mode, the sweep time
may be set to faster values when Option AYX is
installed.
Front Panel
Access:
Sweep, Sweep Time Auto Man
Automatic Sweep Time
[:SENSe]:SWEep:TIME:AUTO OFF|ON|0|1
[:SENSe]:SWEep:TIME:AUTO?
Automatically selects the fastest sweep time for the current settings.
Factory Preset
and *RST:
On
Front Panel
Access:
Sweep, Sweep Time Auto Man
Chapter 5
Language Reference
SENSe Subsystem
5-105