[:SENSe]:SWEep Subsection

Default Unit:
90
Input Port Power Gain
[:SENSe]:POWer[:RF]:GAIN[:STATe] OFF|ON|0|1
[:SENSe]:POWer[:RF]:GAIN[:STATe]?
Turns the internal preamp on or off.
*RST:
Off
Remarks:
This command is only available with Option PA3 installed.
Key access:
AMPTD > Preamp On Off
Sweep Time
[:SENSe]:SWEep:TIME <time>
[:SENSe]:SWEep:TIME?
Specifies the time in which the instrument sweeps the
display.
*RST:
100.2 ms
seconds
Remarks:
A span value of 0 Hz causes the analyzer to enter zero span
mode. In zero span the X- axis represents time rather than
frequency.
Key access:
BW/SWP > Sweep Time Auto Man
Automatic Sweep Time
[:SENSe]:SWEep:TIME:AUTO OFF|ON|0|1
[:SENSe]:SWEep:TIME:AUTO?
Automatically selects the fastest sweep time for the current
settings.
*RST:
On
Key access:
BW/SWP > Sweep Time Auto Man
N9340B Programmer's Guide