Automatic Alignment

:CALibration:AUTO OFF|ON|0|1
:CALibration:AUTO?
Turns the automatic alignment on and off. This is run continuously, at
the end of each sweep.
Factory Preset
and *RST:
Front Panel
Access:

Return to the Default Alignment Data

:CALibration:DATA:DEFault
Initializes the alignment data to the factory defaults.
Front Panel
Access:

Align FM Demodulation

:CALibration:FMDemod
:CALibration:FMDemod?
Performs an alignment of the FM Demodulation board if Option BAA
(FM Demodulation) is installed. The query form of this command
performs the alignment and returns zero if the alignment is successful.
Both this command and front panel access are available only when
NOTE
Option BAA (FM Demodulation) is installed.
Front Panel
Access:
Query the Internal or External Frequency
Reference
:CALibration:FREQuency:REFerence?
This is a query only.
Front Panel
Access:
Chapter 5
On at power-up
System, Alignments, Auto Align, All
System, Alignments, Auto Align, All but RF
System, Alignments, Auto Align, Off
System, Alignments, Load Defaults
System, Alignments, Align Now, FM Demod
none
Language Reference
CALibration Subsystem